Why gas purity so important ?
Gas traps is to remove detrimental impurities from carrier and detector gases. Combination traps are available which remove moisture, oxygen, and/or organics with a single trap. The effectiveness of the traps depends on the initial quality of the gas. Constant exposure of capillary columns to oxygen and moisture, especially at high temperatures, results,in rapid and severe column damage. The use in-line traps for the carrier gas may extend column life and protect the instrument. Any moisture or oxygen introduced into the gas stream due to a leak will be removed by the trap until it expires.
Applied Instruments
GC-MS、TCD、ECD、FID、FPD、PFPD、TSD 、TID、NPD、PND、LC-MS、ICP-MS 、ICP-OES、TOC、Zero-Air
Product Specification

Cart.No

GLEQ-711001

GLEQ-711002

GLEQ-712003

Product discription

Big hydrocarbon trap

Big miosture trap

Big Hydrocarbon & Moisture Trap

Outline gas efficiency

> 99.9999%

Volume

750cc

Absorbing volume

80g (hydrocarbon)

130g(miosture)

40g(hydrocarbon)60g( miosture )

After volume

15ppb (hydrocarbon )

45ppb(miosture)

15ppb (C8 hy),45ppb(miosture)

Max proof pressure

250 psig

Max temperature

100℃

Max flow-rate

25L/min

Apply inert gas type

He, H2, N2, AR, Air

Adapter size

1/8”

Weight

1.5Kg

Life time

13 bottles 40L gases / 2 year

6 bottles 40L gases / 1 year